Developers - Post-market surveillance of medical devices
Reviewed: 14 July 2023 • Last updated: 14 July 2023 • What's new
After your medical device is placed on the UK market, you should monitor the device and report any safety incidents to the MHRA. This is known as post-market surveillance.
Why you should do post-market surveillance
Post-market surveillance makes sure your device is acceptably safe to use for as long as it is in use. If you do not follow the requirements the MHRA may prosecute you. This could result in 6 months in prison or an unlimited fine.
How to do post-market surveillance
Make sure your quality management system is set up appropriately
You should collect and analyse data on the performance and safety of your medical device in a systematic manner. You will use your quality management system to help you do this. Having this system in place is a core part of medical device safety. So, it’s important to make sure the system is set up appropriately.
Your quality management system will likely need:
- processes for collecting information (for example, feedback or complaints from users and audit findings)
- analytical methods to assess collected data and identify reportable incidents
- procedures and systems for investigating the cause of safety concerns and implementing corrective actions
- tools to trace and identify the device in case corrective actions are needed (for example, traceability of a potentially defective device so it can be recalled)
See our guidance on planning your quality management system and the relevant standards you should follow to create it.
Identifying an adverse incident
A safety concern is deemed to be a reportable adverse incident when it meets these 3 criteria:
- an event has occurred
- the device is suspected to be a contributory cause of the event
- the event resulted in, or may have resulted in, death or a serious deterioration in the state of health of a patient or person
Note that an ‘event’ is not limited to use of the device by or for a patient. An ‘event’ includes device testing, examination of supplied information or wider scientific information that could lead or has led to an event.
Not all reportable events result in direct harm from the situation or an intervention. An event is still reportable if no injury was sustained but could occur if the event is repeated.
Reporting an adverse incident to the MHRA
You should notify the MHRA as soon as you become aware that your device may have caused or contributed to a reportable incident. Report individual incidents on the MORE portal.
Each individual report must lead to a final vigilance report, unless you combine the initial and final report into one report. The final report should include your root cause analysis and details of any similar incidents that have occurred, along with the corrective or preventative measures that you took.
You can find guidance on how to analyse the root cause in ISO 14971.
Some adverse incidents can be submitted as part of periodic summary reporting. You would have to arrange details of the timing and content of periodic summary reports on an individual basis with the MHRA.
For more information on the vigilance system and what constitutes a reportable incident, see the MHRA’s guidance on vigilance.
This information is not intended to replace formal statutory guidance regarding legal requirements. For an authoritative view of what regulations require beyond this digest, please see the relevant gov.uk web pages pertaining to the MHRA.
Get more support
To discover how the MHRA can assist you and for contact details, visit our 'Get Support' page.